Add to Cart
High Precision Ultrasonic Thickness Gauge Meter to Measure Plastics / Glass / Ceramics
1. Introduction
ML160 ultrasonic thickness gauge is used to measure the thickness of materials that can enter only one of the sides. this instrument is suitable for many different materials including steel, cast iron, aluminum, red copper, brass, zinc, quartz glass, polyethylene PVC, gray cast iron, ductile iron, iron and other ultrasonic good conductive materials. The dual probe can extend the measurement range to all possible ranges. Sensor models are available special applications, including coarse particle materials and high temperature applications. Two point calibration function; two working modes: single point mode and scanning mode; coupling state indicator showing coupling state; automatic sleep and automatic power off function to save battery life.
2 Functions
1)Suitable for measuring the thickness of any hard material, such as steel, cast iron, aluminum, copper, brass, zinc, quartz glass, polyvinyl chloride, gray cast iron, ductile iron, etc
2)Auto Zero
3)Data storage, up to 20 t files (up to 99 values per file)
4)Automatic Power Off
5)High precision linear compensation circuit (1% H 0.1)
6)Sound speed test (m/s) with 12 storage memory.
7)2.5 MHz for rough surface measurements
3. Specification
Display | 4.5 digits LCD with EL backlight |
Measuring Range | 0.75~300mm (in Steel) |
Sound Velocity Range | 1000~9999 m/s |
ML160 | 0.1/0.01mm |
Accuracy | ±(0.5%Thickness+0.04)mm, depends on materials and conditions |
Units | Metric/Imperial unit selectable |
Power Source | Two “AA” size, 1.5 Volt alkaline batteries. 100 hours typical operating time (EL backlight off |
Communication | USB port |
Outline dimensions | 150×74×32 mm. Weight:245g |
Memory | 20 files (up to 99 values for each file) of stored values |
Mode |
Four measurements readings per second for single point measurement, and ten per second for Scan Mode
|
Appendix: Transducer Selection
Model |
Frequency MHz |
Diameter mm |
Measuring Range | Lower limit | Description |
N02 | 2.5 | 14 | 3.0mm~300.0mm (In Steel) | 20 | for thick, highly attenuating, or highly scattering materials |
N05 | 5 | 10 | 1.2mm~230.0mm (In Steel) | Φ20mm×3.0mm | Normal Measurement |
N05/90° |
5 | 10 | 1.2mm~230.0mm (In Steel) | Φ20mm×3.0mm | Normal Measurement |
N07 | 7 | 6 |
0.75mm~80.0mm (In Steel) |
Φ15mm×2.0mm | For thin pipe wall or small curvature pipe wall measurement |
HT5 | 5 | 12 | 3~200mm (In Steel) | 30 | For high temperature (lower than 300℃) measurement |